Magnetic measurement of the critical current anisotropy in coated conductors

F. Hengstberger, M. Eisterer, H. W. Weber

arXiv:1101.1788 · DOI 10.1088/0953-2048/24/4/045002 · 2011-01-10 · Superconductor Science and Technology

T1theoretical3 citations

Active bibliographic source — not scientific approval

Bibliographic access preserves source history; it does not approve extracted materials or validate reported claims. Review warnings on each occurrence separately.

Abstract

We determine the critical current anisotropy at maximum Lorentz force from hysteresis loops in a vibrating sample magnetometer. To eliminate the signal of spurious variable Lorentz force currents it is sufficient to cut the sample to a specific length, which is calculated from the position dependent sensitivity of the instrument. The procedure increases the resolution of the measurement and the results compare well to transport data on the same sample. As the electric field in magnetisation measurements is lower than in transport experiments the anisotropy at high currents (low temperatures and fields) can be measured without the need of making current contacts or any special sample preparation.