Characterizing cryogenic amplifiers with a matched temperature-variable noise source

Slawomir Simbierowicz, Visa Vesterinen, Joshua Milem, Aleksi Lintunen, Mika Oksanen, Leif Roschier, Leif Grönberg, Juha Hassel, David Gunnarsson, Russell E. Lake

arXiv:2009.03010 · DOI 10.1063/5.0028951 · 2020-09-07 · The Review of scientific instruments

T3experimental43 citations

Active bibliographic source — not scientific approval

Bibliographic access preserves source history; it does not approve extracted materials or validate reported claims. Review warnings on each occurrence separately.

Abstract

We present a cryogenic microwave noise source with a characteristic impedance of 50 $Ω$, which can be installed in a coaxial line of a cryostat. The bath temperature of the noise source is continuously variable between 0.1 K and 5 K without causing significant back-action heating on the sample space. As a proof-of-concept experiment, we perform Y-factor measurements of an amplifier cascade that includes a traveling wave parametric amplifier and a commercial high electron mobility transistor amplifier. We observe system noise temperatures as low as $680^{+20}_{-200}$ mK at 5.7 GHz corresponding to $1.5^{+0.1}_{-0.7}$ excess photons. The system we present has immediate applications in the validation of solid-state qubit readout lines.