← Back to search

Coulomb blockade thermometry based nanocalorimetry

Mari C. Cole, Maximilian T. Pelly, Craig V. Topping, Thomas Reindl, Ulrike Waizmann, Jürgan Weis, Andreas W. Rost

arXiv:2507.11327 · DOI 10.1063/5.0257902 · 2025-07-15 · The Review of scientific instruments

T3experimental1 citations

Active bibliographic source — not scientific approval

Bibliographic access preserves source history; it does not approve extracted materials or validate reported claims. Review warnings on each occurrence separately.

Abstract

Specific heat is a powerful probe offering insights into the entropy and excitation spectrum of the studied material. While it is well established, a key challenge remains the measurements of microcrystals or thin films especially in the sub-Kelvin, high magnetic field regime. Here we present a setup combining the high sensitivity of SiN$_x$ membrane based calorimetry with the absolute accuracy of Coulomb blockade thermometry to realise a nanocalorimeter for such tasks. The magnetic field independent technique of Coulomb blockade thermometry provides an on-platform thermometer combining a primary thermometry mode for in-situ calibration with a fast secondary mode suitable for specific heat measurements. The setup is validated using measurements of a 20 $μ$g sample of Sr$_3$Ru$_2$O$_7$ achieving a resolution on the order of 0.1 nJ/K at 500 mK and an absolute accuracy limited by the determination of the sample's mass. Measurements of CeRh$_2$As$_2$ further highlight the benefits of measuring microcrystals with such a device.

Similar papers