The persistent current switch (PCS) closes the superconducting loop of an MRI magnet. During ramping it is commonly treated as a binary element: with the heater on, the PCS branch is assumed effectively open; after heater turn-off, it is assumed superconducting. Field measurements with an MPS2 service power supply (Digit Systems) on a Siemens MAGNETOM Avanto show that the PCS remains electrically active and that its resistive state can persist after heater turn-off through Joule self-heating. Dedicated maneuvers identified two resistive regimes. With the heater on, the low-voltage resistance is nearly constant; after heater turn-off, self-heating produces a power-law voltage dependence. Both are described by a single effective electrothermal relation, $R_{sw}=C\,(P_h h+V_{node}^2/R_{sw})^a$, where $h$ is the recorded heater state, spanning a 690-fold resistance range. Final-hold data from 17 Siemens magnet sessions confirmed the voltage dependence of effective PCS resistance. Since branch-current relaxation follows $τ=L/R_{sw}$, the required hold time depends on switch state rather than a fixed delay; the median margin after source settling was $3.6\,τ$. The intersection of the PCS characteristic with the external-circuit load line defines the boundary of self-sustained resistive operation. For Avanto, the predicted 0.062 V threshold lay between superconducting recovery at 0.06 V and resistive-state persistence at 0.07 V, and retrospectively classified all 24 archived segments correctly. Including both PCS resistive regimes in the dynamic model reproduced the observed current redistribution: RMSE decreased from 1.449 to 0.154 A, a factor of 9.4. This improvement required explicit representation of heater state.

Source-reported materials — not catalogue approval

FormulaReported Tc (K)Pressure (GPa)Type
NbTi

Archive — visibility unverified

Source-occurrence policy only; no material identity or catalogue acceptance is inferred from the formula.

—Pressure not reportedunknown