← Back to search

Loss tangent measurements of dielectric substrates from 15 K to 300 K with two resonators: investigation into accuracy issues

Janina E. Mazierska, Mohan V. Jacob, Dimitri O. Ledenyov, Jerzy Krupka

arXiv:1210.2230 · DOI 10.1109/APMC.2005.1606805 · 2012-10-08 · 2005 Asia-Pacific Microwave Conference Proceedings

T3experimental6 citations

Active bibliographic source — not scientific approval

Bibliographic access preserves source history; it does not approve extracted materials or validate reported claims. Review warnings on each occurrence separately.

Abstract

The loss tangent of medium, low and very low loss dielectric substrates (including the Rogers RT Duroid 5880 and 6010.2, LaAlO3, (La, Sr)(Al, Ta)O3, MgO and Quartz) was measured at varying temperatures with two TE01δ dielectric resonators to ensure verification of the tests. The accuracy of the measurements has been researched and discussed for split post dielectric resonator (SPDR) in a copper enclosure and a single post dielectric resonator (SuPDR) in a superconducting enclosure in the temperature range from 15K to 300 K.

Similar papers