← Back to search

Scanning microSQUID Force Microscope

C. Veauvy, D. Mailly, K. Hasselbach

arXiv:cond-mat/0110196 · 2001-10-10 · arXiv: Superconductivity

T3

Active bibliographic source — not scientific approval

Bibliographic access preserves source history; it does not approve extracted materials or validate reported claims. Review warnings on each occurrence separately.

Abstract

A novel scanning probe technique is presented: Scanning microSQUID Force microscopy (SSFM). The instrument features independent topographic and magnetic imaging. The SSFM operates in a dilution refrigerator in cryogenic vacuum. Sample and probe can be cooled to 0.45 K. The probe consists of a microSQUID placed at the edge of a silicon chip attached to a quartz tuning fork. A topographic vertical resolution of 0.02 micrometer is demonstrated and magnetic flux as weak as $10^{-3} Φ_{0}$ is resolved with a 1 micrometer diameter microSQUID loop.