Design concepts for an improved integrated scanning SQUID
Nicholas C. Koshnick, John R. Kirtley, Kathryn A. Moler
arXiv:1002.1529 · 2010-02-08 · arXiv: Superconductivity
Active bibliographic source — not scientific approval
Bibliographic access preserves source history; it does not approve extracted materials or validate reported claims. Review warnings on each occurrence separately.
Abstract
In this paper we discuss design concepts for increasing the spatial resolution, improving the sensitivity, and reducing the invasiveness in scanning Superconducting Quantum Interference Device (SQUID) microscope sensors with integrated flux pickup loops. This can be done not only by reducing the ground-rule line widths and spacings, but also by taking advantage of planarization, reducing flux noise through reducing the SQUID inductance, and reducing back-action through dispersive readouts or on-chip filtering.
Similar papers
Spatial resolution and point spread function of high-resolution scanning SQUID microscopy probes
similarity 0.90Jan Ullmann et al. · 2026 · arXiv:2607.17414
Source status unknown — claims are unverified
3D nano-bridge-based SQUID susceptometers for scanning magnetic imaging of quantum materials
similarity 0.90Y. P. Pan et al. · 2019 · arXiv:1903.11262
Source status unknown — claims are unverified
Planar Superconductor-Ferromagnet-Superconductor Josephson Junctions as Scanning-Probe Sensors
similarity 0.89T. Golod et al.
Source status unknown — claims are unverified
Fundamental studies of superconductors using scanning magnetic imaging
similarity 0.88J. R. Kirtley · 2010 · arXiv:1008.3179
Source status unknown — claims are unverified
Noise scaling in SQUID arrays
similarity 0.88Oscar A. Nieves & Karl-H Müller · 2024 · arXiv:2409.19837
Source status unknown — claims are unverified
Optimizing the spin sensitivity of grain boundary junction nanoSQUIDs -- towards detection of small spin systems with single-spin resolution
similarity 0.87Roman Wölbing et al. · 2013 · arXiv:1301.1189
Source status unknown — claims are unverified