Performance of an X-ray single pixel TES microcalorimeter under DC and AC biasing

Luciano Gottardi, Jan van der Kuur, Piet A. J. de Korte, Roland Den Hartog, Bob Dirks, Manuela Popescu, Henk F. C. Hoevers, Marcel Bruijn, Maria Parra Borderias, Y. Takei

arXiv:1604.02536 · DOI 10.1063/1.3292399 · 2016-04-09 · arXiv: Instrumentation and Detectors

T35 citations

Active bibliographic source — not scientific approval

Bibliographic access preserves source history; it does not approve extracted materials or validate reported claims. Review warnings on each occurrence separately.

Abstract

We are developing Frequency Domain Multiplexing (FDM) for the read-out of TES imaging microcalorimeter arrays for future X-ray missions like IXO. In the FDMconfiguration the TES is AC voltage biased at a well defined frequencies (between 0.3 to 10MHz) and acts as an AM modulating element. In this paper we will present a full comparison of the performance of a TES microcalorimeter under DC bias and AC bias at a frequency of 370kHz. In both cases we measured the current-to-voltage characteristics, the complex impedance, the noise, the X-ray responsivity, and energy resolution. The behaviour is very similar in both cases, but deviations in performances are observed for detector working points low in the superconducting transition (R/RN < 0.5). The measured energy resolution at 5.89keV is 2.7eV for DC bias and 3.7eV for AC bias, while the baseline resolution is 2.8eV and 3.3eV, respectively.