Study of the dependency on magnetic field and bias voltage of an AC-biased TES microcalorimeter

Luciano Gottardi, James Adams, Catherine Bailey, Simon Bandler, Marcel Bruijn, James Chervenak, Megan Eckart, Frank Finkbeiner, Roland den Hartog, Henk Hoevers, Richard Kelley, Caroline Kilbourne, Piet de Korte, Jan van der Kuur, Mark Lindeman, Frederick Porter, Jack Sadleir, Steve Smith

arXiv:1604.02595 · DOI 10.1007/s10909-012-0494-x · 2016-04-09 · Journal of Low Temperature Physics

T3experimental15 citations

Active bibliographic source — not scientific approval

Bibliographic access preserves source history; it does not approve extracted materials or validate reported claims. Review warnings on each occurrence separately.

Abstract

At SRON we are studying the performance of a Goddard Space Flight Center single pixel TES microcalorimeter operated in an AC bias configuration. For x-ray photons at 6 keV the pixel shows an x-ray energy resolution $dE_{FWHM} = 3.7\mathrm{eV}$, which is about a factor 2 worse than the energy resolution observed in an identical DC-biased pixel. In order to better understand the reasons for this discrepancy we characterised the detector as a function of temperature, bias working point and applied perpendicular magnetic field. A strong periodic dependency of the detector noise on the TES AC bias voltage is measured. We discuss the results in the framework of the recently observed weak-link behaviour of a TES microcalorimeter