Cryogen-free variable temperature scanning SQUID microscope

Logan Bishop-Van Horn, Zheng Cui, John R. Kirtley, Kathryn A. Moler

arXiv:1812.03215 · DOI 10.1063/1.5085008 · 2018-12-07 · The Review of scientific instruments

T1experimental18 citations

Active bibliographic source — not scientific approval

Bibliographic access preserves source history; it does not approve extracted materials or validate reported claims. Review warnings on each occurrence separately.

Abstract

Scanning Superconducting QUantum Interference Device (SQUID) microscopy is a powerful tool for imaging local magnetic properties of materials and devices, but it requires a low-vibration cryogenic environment, traditionally achieved by thermal contact with a bath of liquid helium or the mixing chamber of a "wet" dilution refrigerator. We mount a SQUID microscope on the 3 K plate of a Bluefors cryocooler and characterize its vibration spectrum by measuring SQUID noise in a region of sharp flux gradient. By implementing passive vibration isolation, we reduce relative sensor-sample vibrations to 20 nm in-plane and 15 nm out-of-plane. A variable-temperature sample stage that is thermally isolated from the SQUID sensor enables measurement at sample temperatures from 2.8 K to 110 K. We demonstrate these advances by imaging inhomogeneous diamagnetic susceptibility and vortex pinning in optimally-doped YBCO above 90 K.

Source-reported materials — not catalogue approval

FormulaReported Tc (K)Pressure (GPa)Type
YBa2Cu3O7-δ

Archive — visibility unverified

Source-occurrence policy only; no material identity or catalogue acceptance is inferred from the formula.

—Pressure not reportedunknown

Similar papers