Precise magnetization measurements down to 500 mK using a miniature 3He cryostat and a closed-cycle 3He gas handling system installed in a SQUID magnetometer without continuous-cooling functionality

Kazutoshi Shimamura, Hiroki Wajima, Hayato Makino, Satoshi Abe, Yoshinori Haga, Yoshiaki Sato, Tatsuya Kawae, Yasuo Yoshida

arXiv:2202.09067 · DOI 10.35848/1347-4065/ac5bb0 · 2022-02-18

Active bibliographic source — not scientific approval

Bibliographic access preserves source history; it does not approve extracted materials or validate reported claims. Review warnings on each occurrence separately.

Abstract

We have conducted precise magnetization measurements down to 0.5 K with a miniature 3He cryostat and a closed-cycle 3He gas handling system for a commercial superconducting quantum interference device magnetometer [Magnetic Property Measurement System (Quantum Design)]. The gas handling system contains two sorption pumps filled with granular charcoals. We pressurize 3He gas up to ambient pressure for the liquification at 3 K and then pump the vaper for cooling. The lowest sample temperature is ~ 0.5 K and it can persist for 34 hours. We demonstrate the performance of the system by observing the Meissner effect of aluminum below the superconducting transition temperature ~ 1 K. We also measured the magnetization curve of the heavy fermion superconductor CeCoIn5 resulting in successful observation of the lower critical field at 0.5 K.

Source-reported materials — not catalogue approval

FormulaReported Tc (K)Pressure (GPa)Type
Al

Archive — visibility unverified

Source-occurrence policy only; no material identity or catalogue acceptance is inferred from the formula.

1Pressure not reportedunknown
CeCoIn5

Archive — visibility unverified

Source-occurrence policy only; no material identity or catalogue acceptance is inferred from the formula.

—Pressure not reported—

Similar papers