Beamline Spectroscopy of Integrated Circuits With Hard X-ray Transition Edge Sensors at the Advanced Photon Source

T. Guruswamy, L. Gades, A. Miceli, U. Patel, O. Quaranta

arXiv:2012.11680 · DOI 10.1109/TASC.2021.3067246 · 2020-12-21 · IEEE Transactions on Applied Superconductivity

T1theoretical5 citations

Active bibliographic source — not scientific approval

Bibliographic access preserves source history; it does not approve extracted materials or validate reported claims. Review warnings on each occurrence separately.

Abstract

At Argonne National Laboratory, we are developing hard X-ray (2 to 20 keV) Transition Edge Sensor (TES) arrays for beamline science. The significantly improved energy resolution provided by superconducting detectors compared to semiconductor-based energy-dispersive detectors, but with better collection efficiency than wavelength-dispersive instruments, will enable greatly improved X-ray emission and absorption spectroscopic measurements. A prototype instrument with 24 microwave-frequency multiplexed pixels is now in testing at the Advanced Photon Source (APS) 1-BM beamline. Initial measurements show an energy resolution ten times better (150 eV compared to < 15 eV) than the silicon-drift detectors currently available to APS beamline users, and in particular demonstrate the ability to resolve closely-spaced emission lines in samples containing multiple transition metal elements, such as integrated circuits. Comparing fluorescence spectra of integrated circuits measured with our TESs at the beamline to those measured with silicon detectors, we find emission lines and elements largely hidden (e.g. Hf alongside Cu) from a semiconductor-based detector but well resolved by a TES. This directly shows the strengths of TES-based instruments in fluorescence mapping.

Source-reported materials — not catalogue approval

FormulaReported Tc (K)Pressure (GPa)Type
MoCu

Archive — visibility unverified

Source-occurrence policy only; no material identity or catalogue acceptance is inferred from the formula.

0.0875Pressure not reportedunknown

Similar papers