A high-speed, high-resolution Transition Edge Sensor spectrometer for soft X-rays at the Advanced Photon Source
Orlando Quaranta, Don Jensen, Kelsey Morgan, Joel C. Weber, Jessica L. McChesney, Hao Zheng, Tejas Guruswamy, Jonathan Baldwin, Ben Mates, Nathan Ortiz, Johnathon Gard, Doug Bennet, Dan Schmidt, Lisa Gades, Antonino Miceli
arXiv:2501.01883 · 2025-01-03
Active bibliographic source — not scientific approval
Bibliographic access preserves source history; it does not approve extracted materials or validate reported claims. Review warnings on each occurrence separately.
Abstract
This project explores the design and development of a transition edge sensor (TES) spectrometer for resonant soft X- ray scattering (RSXS) measurements developed in collaboration between Argonne National Laboratory (ANL) and the National Institute of Standards and Technology (NIST). Soft X-ray scattering is a powerful technique for studying the electronic and magnetic properties of materials on a microscopic level. However, the lack of high-performance soft X-ray spectrometers has limited the potential of this technique. TES spectrometers have the potential to overcome these limitations due to their high energy resolution, high efficiency, and broad energy range. This project aims to optimize the design of a TES spectrometer for RSXS measurements and more generally soft X-ray spectroscopy at the Advanced Photon Source (APS) 29-ID, leading to improved understanding of advanced materials. We will present a detailed description of the instrument design and implementation. The spectrometer consists of a large array of approximately 250 high-speed and high-resolution pixels. The pixels have saturation energies of approximately 1 keV, sub-ms pulse duration and energy resolution of approximately 1 eV. The array is read out using microwave multiplexing chips with MHz bandwidth per channel, enabling efficient data throughput. To facilitate measurement of samples in situ under ultra-high vacuum conditions at the beamline, the spectrometer is integrated with an approximately 1 m long snout.
Similar papers
Beamline Spectroscopy of Integrated Circuits With Hard X-ray Transition Edge Sensors at the Advanced Photon Source
similarity 0.87T. Guruswamy et al. · 2020 · arXiv:2012.11680
Source status unknown — claims are unverified
Hard X-ray Fluorescence measurements with TESs at the Advanced Photon Source
similarity 0.87Tejas Guruswamy et al. · 2019 · arXiv:1909.05919
Source status unknown — claims are unverified
Characterization of Silicon-Membrane TES Microcalorimeters for Large-Format X-ray Spectrometers with Integrated Microwave SQUID Readout
similarity 0.86Avirup Roy et al. · 2026 · arXiv:2601.09629
Source status unknown — claims are unverified
Proposal for a GHz count rate near-IR single-photon detector based on a nanoscale superconducting transition edge sensor
similarity 0.84Daniel F. Santavicca et al. · 2012 · arXiv:1202.4722
Source status unknown — claims are unverified
Proximity Effects and Nonequilibrium Superconductivity in Transition-Edge Sensors
similarity 0.84John E. Sadleir et al. · 2011 · arXiv:1108.4632
Source status unknown — claims are unverified