Hard X-ray Fluorescence measurements with TESs at the Advanced Photon Source

Tejas Guruswamy, Lisa M. Gades, Antonino Miceli, Umeshkumar M. Patel, John T. Weizeorick, Orlando Quaranta

arXiv:1909.05919 · DOI 10.1088/1742-6596/1559/1/012018 · 2019-09-12

Active bibliographic source — not scientific approval

Bibliographic access preserves source history; it does not approve extracted materials or validate reported claims. Review warnings on each occurrence separately.

Abstract

Transition Edge Sensor (TES) spectrometers for hard X-ray beamline science will enable improved X-ray emission and absorption spectroscopy in the information-rich 2 to 20 keV energy range. We are building a TES-based instrument for the Advanced Photon Source (APS) synchrotron, to be made available to beamline users. 24-pixel prototype arrays have recently been fabricated and tested. The first spectroscopy measurements using these arrays are promising, with a best single-pixel energy resolution of 11.2 eV and saturation energy > 20 keV. We present a series of recent X-ray Fluorescence measurements involving transition metal elements and multi-element samples with closely spaced emission lines, in particular a Cu-Ni-Co thin film and a foil of Cu and Hf. The TES-measured spectra are directly compared to spectra measured with silicon drift detectors at an APS beamline, demonstrating the improved X-ray science made possible by TES spectrometers.