← Back to search

Development of TES-based detectors array for the X-ray Integral Field Unit (X-IFU) on the future x-ray observatory ATHENA

Luciano Gottardi, Hiroki Akamatsu, Didier Barret, Marcel P. Bruijn, Roland H. den Hartog, Jan-Willem den Herder, Henk F. C. Hoevers, Mikko Kiviranta, Jan van der Kuur, Anton J. van der Linden, Brian D. Jackson, Madu Jambunathan, Marcel L. Ridder

arXiv:1604.00553 · DOI 10.1117/12.2056920 · 2016-04-02

T3review11 citations

Active bibliographic source — not scientific approval

Bibliographic access preserves source history; it does not approve extracted materials or validate reported claims. Review warnings on each occurrence separately.

Abstract

We are developing transition-edge sensor (TES)-based microcalorimeters for the X-ray Integral Field Unit (XIFU) of the future European X-Ray Observatory Athena. The microcalorimeters are based on TiAu TESs coupled to 250μm squared, AuBi absorbers. We designed and fabricated devices with different contact geometries between the absorber and the TES to optimise the detector performance and with different wiring topology to mitigate the self-magnetic field. The design is tailored to optimise the performance under Frequency Domain Multiplexing. In this paper we review the main design feature of the pixels array and we report on the performance of the 18 channels, 2-5MHz frequency domain multiplexer that will be used to characterised the detector array.